Chapter 10 Operating Instructions
CLV490 Bar Code Scanner
10-62 © SICK AG · Division Auto Ident · Germany · All rights reserved 8009993/RA61/2007-07-20
Appendix
5. Connect the pulse photoelectric switch to the CDB420 or CDM490 connection module
(see Chapter 5.5.7 Connecting the "Sensor" switching input, Page 5-22).
6. Turn on the supply voltage to the CLV490.
10.10.6 Parameterize the CLV490 with the "CLV-Setup" program
1. Connect the PC to the CLV490’s auxiliary interface (see Chapter 5.5.6 Connecting the
PC, Page 5-21)
2. Start Windows and the "CLV-Setup" program on the PC.
3. Invoke the CLV490’s default setting in the "CLV-Setup" program
(see Chapter 6.1 Overview of steps for starting up the CLV490, Page 6-1).
4. Parameterize the CLV490 with the help of Chapter 6.4.4 Guide to parameterization
menu, Page 6-9 and the tab cards by following the steps described.
5. After completing parameter value alterations, download them to the CLV490!
(see chapter Chapter 6.4.1 Configuring the CLV490 via the user interface of CLV-Set-
up, Page 6-5).
1. Setting up the autofocus function (Part 1): R
EADING CONFIGURATION tab
¾ Set the following parameters:
Scanning frequency: 800 Hz
Start/stop ratio: auto
Focus control: autofocus
Autofocus mode: differential background
¾ Click on the
AUTOFOCUS PARAMETERS button.
The
AUTOFOCUS PARAMETERS tab then opens (Fig. 10-46).
¾ Set the following parameters:
Preference of focus distance: Fix focus position: 1,200 mm (47.3 in)
(approx. 100 mm (3.9 in) above the packet with the smallest height)
Fig. 10-45: Parameterization example: settings on the "Reading Configuration" tab